Application |
---|
Application | for Pt100 and Pt1000 sensors |
Electronic data |
---|
Operating voltage [V] | 18…32 DC, ("supply class 2" according to cULus) |
Current consumption [mA] | <50 |
Protection class | III |
Reverse polarity protection | yes |
Power-on delay time [s] | 1 |
Built-in watchdog | yes |
Inputs/outputs |
---|
Number of inputs and outputs | Number of digital outputs: 1, Number of analog outputs: 1 |
Outputs |
---|
Total number of outputs | 2 |
Output signal | Switching signal, analogue signal , IO-Link, (configurable) |
Electrical version | PNP/NPN |
Number of digital outputs | 1 |
Output function | n normally open / normally closed, (parameterizable) |
Max. voltage drop of switching output DC [V] | 2.5 |
DC load current of switching output DC [mA] | 250 |
Number of analog outputs | 1 |
Analog current output [mA] | 4…20 |
Maximum load [?] | 500 |
Analog output according to voltage [V] | 0…10 |
Min. load resistance [?] | 2000 |
Short circuit protection | yes |
Type of short-circuit protection | clocked |
Overcurrent protection | yes |
Measuring/setting range |
---|
Measuring range | |
Factory setting | -40…300 °C / -40…572 °F |
Setting point SP | -99.8…600 °C | -147.6…1112 °F |
|
Reset point rP | -100…599.8 °C | -148…1111.6 ° F |
|
Aanalog trigger point | |
Analog endpoint | |
In steps | |
Resolution |
---|
Resolution of switching output [K] | 0.1 |
Analog output resolution [K] | current output: MS / 4096, voltage output: MS / 3561 |
Display Resolution [K] | 0.1 |
Accuracy/Uncertainty |
---|
Switch point error [K] | ± 0.3 + (± 0.1 % MS) |
Analog output accuracy [K] | ± 0.3 + (± 0.1 % MS) |
Display Accuracy [K] | ± 0.3 + (± 0.1 % MS) |
Temperature coefficient [% of measuring range / 10 K] | 0.1, (When deviating from standard conditions 25 ± 5 °C) |
Response time |
---|
Response time [ms] | 390 |
Max. analog output response time [ms] | 390 |
Software / Programming |
---|
Parameter selection | Hysteresis / Window, Normally Open / Normally Closed, On / Off Delay, Damping, Display, Current / Voltage Output |
Interfaces | >
---|
Communication interface | IO-Link |
Transmission method | COM2 (38, 4 KBaud) |
IO-Link test | 1.1 |
Operating conditions |
---|
Ambient temperature [°C] | -25…80 |
Storage temperature [°C] | -40…100 |
Degree of protection | IP 67 |
Tests / approvals |
---|
EMC | DIN EN 61000-6-3 | | DIN EN 61000-6-2 | |
|
Impact resistance | DIN IEC 68-2-27 | 50 g (11 ms) |
|
Vibration resistance | DIN EN 60068- 2-6 | 20 g (10…2000 Hz) |
|
MTTF [ years] | 233 |
UL Certification | UL Registration Number | K011 |
|
Mechanical data |
---|
Weight [g] | 200 |
Material | stainless steel steel (1.4404 / 316L), stainless steel steel (1.4301/304), PBT+PC-GF30, PBT-GF20, PC |
Displays / Controls |
---|
Display | Display | 2 x LED, green | Output status | LED, yellow | Measured values | alphanumeric display, red/green 4-digit |
|
Notes |
---|
Notes | MS = measuring interval setting |
|
Pack quantity | 1 piece |
electrical connection |
---|
Connection | Connector: 1 x M12, Contacts: gold-plated |